Performance characteristics conductivity
Conductivity : | 14 selectable ranges |
Min. measurable conductivity : | 50 µS/cm |
Max. measuring span : | 1000 mS/cm |
Min. measuring span : | 500 µS/cm |
Max. measuring error : |
|
Reference temperature : | 25 °C , adjustable |
Repeatability : | < 0.5 % FSR ,> 1 mS/cm |
Compensated temperature range : | -20 … 150 °C |
Temperature compensation : | 0.0 … 5.0 % FSR/K , adjustable |
Step response time : | ≤ 2.0 s |
Sample time : | ≤ 0.4 s |
Performance characteristics concentration
Concentration : | 4 factory set media |
Customer defined media : | Customer defined (30 point lookup table) |
Performance characteristics temperature
Temperature : | Free programmable range |
Measuring range : | -20 … 150 °C |
Thermal response time, T90 : | ≤ 15 s |
Max. measuring error : | ± 0.4 K |
Process conditions
Process temperature : |
|
Process pressure : | ≤ 25 bar |
Process connection
Connection variants : | G 1 A hygienic |
Wetted parts material : | PEEK Natura |
Surface roughness wetted parts : | Ra ≤ 0.8 µm |
HART® interface
Properties : |
|
Protocol : | HCF standard, Rev.7 |
IO-Link interface
IO-Link version : | 1.1 |
Device profile : | Smart Sensor Profile |
Baud rate : | 38,4 kbaud (COM2) |
Cycle time : | ≥ 8,4 ms |
Process data length : | 128 bit |
SIO-mode : | Yes |
Process data (cyclic) : |
|
Ambient conditions
Degree of protection (EN 60529) : |
|
Cable (AFI5)
Cable lengths : |
|
Material : | PUR |
Temperature : | -40 … 80 °C |
Compliance and approvals
EMC : | EN 61326-1 |
Hygiene : |
|
Safety : | cULus listed, E491206 |
Download Broucher : | AFI5 |
CombiLyz AFI5
- Separated sensor
- Ideal for cramped spaces and strong vibrations
- All wetted parts in PEEK
- Compact, food-safe, hygienic design
- 3-A sanitary standards, FDA-compliant, EHEDG-certified
- Available with HART® communication protocol
- Available with IO-Link communication interface